High Technology, Inc.
109 Production Road,
Walpole MA 02081 USA
Document Number
OM-E-FC-360-30
Revision Level
Rev. 3
Prepared by
J.S. Bolio
Approved by
S. Titov
Effective Date
8/15/2016
Operator’s Manual: HTI BioChem FC-360
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
Page 147 of 222
Washer purge
The washing of the first 10 cuvettes is carried out during this cycle, which allows the filling of all tubing of
the washer circuit and the elimination of bubbles.
This cycle is carried out:
-
after adding washing solution to the container
-
every time it is necessary to purge tubing
Initialize
By clicking this command an individual initialization of each and every module will be executed, following
a certain order: photometer, probe, probe arm, SR tray, diluter, washer and cuvettes tray.
This cycle is performed:
-
At the beginning of daily work
-
Automatically after an auto-stop sequence
ISE (Optional)
Through this menu different ISE maintenance options can be run.
Figure 12-50