Operator’s Manual: HTI BioChem FC-360
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
Page 63 of 222
7. Press
Verify Selected
or
Do not verify
. If the option
Do Not Verify
is selected, the methods that are
checked will not be verified.
Warning: This is the last screen where processing of requested samples can be cancelled.
8. Press
Start
on the screen to put the samples in the tray (Figure 7-31)
Warning: It is not possible to cancel processing of selected samples in this step.
Figure 7-31
Samples to Tray (Barcode) (Optional)
Processing of samples with barcodes can be initialized from the
Samples to Tray (Barcode)
submenu or
from the
Samples to Tray (Barcode)
icon when the instrument is installed with the optional barcode
reader (Figure 7-32)
Figure 7-32
To begin processing samples, follow the next steps:
1. Add the daily work list of samples (either manually or imported from a pending file or a .csv)
Warning:
First methods should be calibrated and controlled, once the values for reagent blank, calibrator
absorbance and quality controls are validated, then proceed to process
patients’ samples.
2. Press
Samples to Tray
(Barcode)
The analyzer will scan every sample position in the SR tray until it finds all the samples added in the work
list, meaning, if the samples in the work list are in the first ten positions then the analyzer will find the samples