Operator’s Manual: HTI BioChem FC-360
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
Page 209 of 222
Reference amplifier failure.
Contact Technical Support.
Filter missing
Contact Technical Support.
Incorrect positioning of a Beam Filter. Contact Technical Support.
(0330) High energy Reference
Beam
Reaction tray lid not placed.
Place the lid of the reaction tray.
Amplifiers disconnected.
Contact Technical Support.
Filter missing
Contact Technical Support.
Incorrect positioning of a Beam Filter. Contact Technical Support.
(0331) High energy Sample
Beam
Lamp failure
Contact Technical Support.
Incorrect positioning of a Beam Filter. Contact Technical Support.
Defective filter
Contact Technical Support.
(0332) Saturated Sample
Lamp failure
Contact Technical Support.
and Reference Amplifiers
Incorrect positioning of a Beam Filter. Contact Technical Support.
Defective filter
Contact Technical Support.
PHOTOMETER CALIBRATION
Message ID Number and
Description
Probable cause
Possible Solution
(0351) Photometer Calibration
error (no filters programmed in
EEPROM)
Filter ID values set on the EEPROM
Contact Technical Support.
(0352) Photometer Calibration
error (Filter Wheel Controller
communication)
Possible failure in filter wheel controller. Contact Technical Support.
(0353) Photometer Calibration
error (Amplifier Controller
communication)
Possible failure in amplifier IC controller. Contact Technical Support.
(0354) Photometer Calibration
error
Calibration logic error.
Contact Technical Support.
(0355) Photometer Calibration
error (low energy in Sample
Beam)
Scratched cuvette or not aligned.
Replace the cuvette or try to calibrate
another cuvette. (Anyway, it is
recommended to replace defective
cuvettes)