Operator’s Manual: HTI BioChem FC-360
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
Page 111 of 222
Figure 11-5
The graph obtained is a Levy-Jennings Plot where the results obtained for each point are indicated. The
CV% and the Average for the data are also reported. In the plot, different lines can be observed, which
show the values corresponding to the average and +/- SD per line (Figure 11-4). The plotted data can be
edited: place the cursor on the appropriate box, enter the new value and press
Enter
on the keyboard. By
pressing
Calculate
, the statistical parameters will be recalculated and a new graph will be obtained.
Warning: Data can be temporarily edited, but will not be saved by the program.
Levy-Jennings of Controls
Levy-Jennings plots can be made to represent values obtained in the same control for a certain analyte.
To do this, data must be previously ordered in the
CON Results
Tab, performing an advanced search by
Control and Method, so that all results for the same method are shown together.
The following steps must be carried out:
1. Select the Process Date from/to which the search will be performed.
2. Perform an advanced search of the data of a level or kind of control and a certain method:
Example: CONTROL N, Method: ALBU-DICO (Figure 11-6).
3. Select the data obtained, press
Statistics Menu
Controls Levy-Jennings Plot
or press the
shortcut
Levy-Jennings Plot
icon.