Operator’s Manual: HTI BioChem FC-360
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
Page 131 of 222
Validations
By clicking on
Maintenance
and selecting
Validations
, some validation tests are displayed which are
used to control how the components of the instrument are working (Figure 12-24). The spectrophotometer,
the diluter and the washer are controlled through these tests. Each user can create an internal quality
control program of their Autoanalyzer by using these tests.
Warning: A Special Washing cycle with Sodium Hydroxide 0.2 N is recommended before making
the Validation Tests (see Chapter 7 point: 7.7)
Figure 12-25