Figure 15-13
Replacing the fuses
1. Turn off the instrument
2. Remove the feed cable from the line filter
3. Remove the lid of the fuse-box located in the upper part of the line filter
4. Remove the burnt fuses and replace them with new ones
5. Replace the lid of the fuse box
6. Reconnect the feed cable to the line filter
7. Turn on the instrument
Special Wash
1. Fill a single mouth reagent container with special washing solution (see chapter 16 point 16.3
and 16.4)
2. Place the container in the reagent tray
3. In the Wash cuvettes screen, select Special Wash and enter in Reagent Pos. the position of
the container
4. Press Wash
5. Once the special wash is finished, the cuvettes are ready to be used as usual
Replacing O-rings of piston pump
Contact Technical Support
Replacing filters of washing system
Contact Technical Support