Operator’s Manual: HTI BioChem FC-360
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
Page 185 of 222
Cuvettes Replacement (see item 15.5.3.8)
15.6.3.1
Interferential filters replacement
a. Turn off the analyzer, remove the lids and the main cover of the autoanalyzer to have access
to the photometer, located on the right back side.
b.
Remove the photometer’s dissipater.
c. Spot the interferential filter to be replaced and manually tu
rn the filter’s wheel clockwise to find
and to easily remove the interferential filter holder from the filter wheel.
d. Place the holder with the new filter in the filter wheel.
Figure 15-6
e. Place back the lids and main cover, switch on the equipment.
f. Run a calibration of the photometer and all the cuvettes
Maintenance
Instrument
Calibration
select
Calibrate photometer
select
Calibrate cuvettes
First cuvette 1
Last cuvette 100
select the position of the solution to use for the calibration ex:
Use
container
Calibrate
Sample / Reagent probe replacement
a. Keep the equipment off, remove the hood of the probe arm.
b. Disconnect the Teflon tubing from the hydraulic connector.