3. Check that the actual positioning of samples on the tray corresponds to what appears on the
Samples
Position
screen.
Sample positions can be modified by using the drop down window in the Position column.
Figure 7-34
If the same position number is used by mistake for two different samples in the same run, an error
message will be displayed requesting to correct the repeated positions (Figure 7-29).
Figure 7-35
4. Select the execution options: by
Selection,
by
Patients
or by
Reagents:
Selection
: This is the fastest option. The software organizes and establishes the work sequence, ordering
of the methods and samples, optimizing processing speed.
Patients
: Processes samples in the order in which they were requested.
Reagents
: All tests of the same method are run before it continues with the tests for the next method. The
user cannot select the methods’ processing order. The software establishes the optimal processing order.
Processing is grouped as
New Group
by default.
5. Press
Send
to continue or
Cancel send
if the operation is not carried out.
6. Select the methods for reagent verification on the
Reagents Verification
screen (Figure 7-30)
After sending samples to be processed, the
Reagent Verification
screen is displayed. In this screen, all
methods where the verification time (configured in the
Verification Time
field) has elapsed are indicated