Operator’s Manual: HTI BioChem FC-360
Page 64 of 217
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
and stop the scanning process after finishing in position 10. If a barcode can’t be read by the analyzer, the
sample corresponding to that barcode won’t be found, therefor the analyzer will continue scanning until it
reaches position 60.
Figure 7-33
Figure 7-34
Once the detecting process is finished, the window for Sample Positions will appear. Verify that the positions
shown in the list match the ones in the SR tray.
NOTE: Those samples with barcodes
that can’t be read by the analyzer will remain in the work list
(Entered Samples), not being sent to tray, the rest of the samples whose barcodes were read OK will
be moved to tray. The samples that remain in the work list can be sent to tray manually or, previously
correcting the barcode (positioning correctly, label, etc), using the barcode reader.
If a new list of samples is scanned in positions that still have pending tests the following message will show:
Figure 7-35
By pressing OK Position 2 will be
scanned again until the original tube is detected. The analyzer won’t allow
to move on until the original tube is placed in its position.