Operator’s Manual: HTI BioChem FC-360
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
Page 91 of 222
C =
Concentration
Ai = A
I
- A
I
’
Af = A
F
- A
F
’
C = (Af
– Ai) • F
NOTE: Always press the Modify and OK buttons if any change is made in the method
configuration. If no parameter is changed, exit by pressing the OK button.
If this is not done, the methods will not be programmed correctly and the results may be
affected.
Advanced tab
Warning: Changes should only be made in this tab under supervision or only by trained
personnel in advanced options.
Figure 8-14
The purpose of modifying this part of the Methods Settings is to improve the performance of a certain
reagent
Within these options are found:
Initial Air Gap:
A volume of air drawn in the diluter. It is used to minimize contact between the reagent
and the liquid inside the probe. The default value is 0 µl and it is recommended not to exceed a value of
20 µl. 2 µl are recommended.
Initial Gap Speed:
The speed with which the diluter draws in the initial air gap. The default value is 500
1
/
6
µl / second.