(0321) The time is not enough
for thermal stabilization. Please
repeat the test after some
minutes
A reading was performed before
reaching the time of thermal
stabilization.
Wait .(it is recommended at least 10
minutes)
(0322) Reaction Tray
uncovered. Please place the
cover
It has been intended to carry out a
reading operation without placing the
cover of the reaction tray.
Place the cover of the reaction tray.
Light sensor failure.
Contact Technical Support.
(0323) Lamp error (burned,
unconnected or power failure)
Lamp burned out.
Replace the lamp.
Lamp disconnected or without any
supplies.
Contact Technical Support.
(0324) Lamp powered off
The lamp turned off and the analyzer
attempts to read.
Contact Technical Support.
(0325) Internal CRC error
(Amplifier)
Communication error between
amplifier controller and Main
controller.
Contact Technical Support.
(0326) Low energy in
Lack of calibration.
Run photometer calibration.
Reference Beam. Calibrate
Photometer
Exit Lens of the reference channel
Contact Technical Support.
Defective Lamp
Contact Technical Support.
Defective Beam Filter
Contact Technical Support.
(0327) Low energy in Sample
Beam. Check Sample
absorbance
Lack of calibration.
Run photometer calibration.
Scratch Cuvette or not aligned.
Contact Technical Support.
Exit Lens of the reference channel
Contact Technical Support.
Defective Lamp
Contact Technical Support.
Defective Beam Filter
Contact Technical Support.
(0328) Saturated Reference
Amplifier
Reaction tray uncover
Place the cover of the reaction tray.
Reference amplifier disconnected.
Contact Technical Support.
Reference amplifier failure.
Contact Technical Support.
Filter missing.
Contact Technical Support.
Incorrect positioning of a Beam Filter. Contact Technical Support.
(0329) Saturated Sample
Amplifier
Reaction tray uncovered
Place the cover of the reaction tray.
Reference amplifier disconnected
Contact Technical Support.