Operator’s Manual: HTI BioChem FC-360
Page 52 of 217
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
7.1.1.2 Add Calibrator
Select
Calibrator
on the
Add Process
screen. All calibrators configured according to Chapter 9 point
9.1 will be displayed. Select the desired calibrator on the left column. The methods to be calibrated with
the selected calibrator will appear in the right column since they have a configured concentration (Figure
7-8). Select the methods to be calibrated and press
Add
.
Each calibrator that has been added will appear on the
Operations tab
with a process number
indicating the time when the request was made and a symbol representing a calibrator
.
Figure 7-8
7.1.1.3 Add Control
Select
Control
on the
Add Process
screen. All controls configured according to Chapter 10, section
10.1 will be displayed. Select the desired control on the left column. The methods for the selected control
will appear in the right column since they have minimum and maximum configured concentrations
(Figure 7-9). Select the methods to be run and press
Add
.
The requested control can be selected as
Stat
.
Each control that has been added will appear on the
Operations tab
with a process number indicating
the time the request was made and a symbol representing a Control
If the option
Stat
is selected, this symbol will appear in red on the requested samples list.
Control
Calibrator