Operator’s Manual: HTI BioChem FC-360
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
Page 165 of 222
Cleaning solution should be used at least once a day at the end of the work day to minimize protein build-
up in the fluid lines and electrodes.
Fluid Management
The sample is aspirated by the analyzer from the sample tube and dispensed into the sample entry port on
top of the ISE Module. The sample is then positioned in front of the electrodes for measurement.
Four solutions are required to operate the ISE Module.
1. Calibrant A is used in both two-point and single-point calibrations for serum sample analysis.
Calibrant A is pumped into the sample entry port by the Calibrant A pump and then positioned
in front of the electrodes by the waste pump. Calibrant A solution is also used for Pump and
Bubble Calibration.
2. Calibrant B is used in two-point and single-point calibrations for urine sample analysis.
Calibrant B is pumped into the sample entry port by the Calibrant B pump and then positioned
in front of the electrodes by the waste pump.
3. Cleaning Solution is used once a day to prevent protein buildup on the electrodes and fluid
path. It must be used more frequently if the ISE Module performs greater than 50 sample
measurements per day. 100 μL of cleaning solution must be aspirated by the host analyzer
from a sample cup on the host analyzer and dispensed into the sample entry port. The
sample cup must be covered to eliminate evaporation.
NOTE: Medica pepsin/HCl cleaning solution must be prepared every four weeks and
stored at 4º C.
4. Medica Urine Diluent. Urine samples must be diluted to perform urine measurement: 1 part
urine sample to 9 parts urine diluent. The diluted specimen must be thoroughly mixed before
aspirating a sample.
Sample type
Sample: Serum, Plasma, or Urine (Urine requires dilution)
Minimum Sa
mple Size: 70 μL serum; 140 μL diluted urine
When measuring urine, the sample must be accurately diluted (1 part sample to 9 parts Medica urine
diluent). The dilution must be performed before the sample is dispensed into the sample entry port.
Warning: Sam
ple volumes should never exceed 150 μL. Volumes greater than 150 μL will result in
the sample mixing with the Calibrant B in the inlet tube.