If single electrode is flagged
Replace the electrode and recalibrate.
If multiple electrodes are flagged
May occur when new electrode or
Reagent Kit is installed.
Purge the Calibrant A and recalibrate
the ISE module.
If multiple electrodes are flagged
Reference electrode.
Replace reference electrode and
recalibrate.
(0709) Noise in Urine
If single electrode is flagged
Replace problem electrode and
recalibrate.
If multiple electrodes are flagged
Replace reference electrode and
recalibrate.
Electrical noise spike from
environmental source.
Contact Technical Support.
Component failure on ISE Module
board.
Contact Technical Support
(0710) Noise in Calibrant B
If single electrodes is flagged
Replace problem electrode and
recalibrate.
If multiple electrodes are flagged
Replace reference electrode and
recalibrate.
Electrical noise spike from
environmental source.
Contact Technical Support.
Component failure on ISE Module
board.
Contact Technical Support
(0711) Drift in Urine
If single electrode is flagged: May
occur when new electrode or
Calibrant B is installed.
Purge the Calibrant B and recalibrate
the ISE module. If the electrode is new it
may initially drift as it rehydrates over
the course of 15 minutes.
If single electrode is flagged
Replace the electrode and recalibrate.
If multiple electrodes are flagged:
May occur when new electrode or
Reagent Kit is installed.
Purge the Calibrant A and recalibrate
the ISE module.
If multiple electrodes are flagged:
Reference electrode.
Replace reference electrode and
recalibrate.
(0712) Out of range in Urine
Samples results are out of range
limits:
Urine
Urine diluted must be used.
Check quality of sample.