Ensure that all electrodes and O-rings are
properly installed.
Ensure tubing between reagent kit and
sample entry port is connected properly.
Replace tubing between reagent kit and
sample entry port.
Reagent low or out.
Fibrin or salt is plugging
the
Use Cleaning procedure.
Electrode flow path.
Remove electrodes and clean or replace
electrode with plugged flow path. Reinstall
electrodes and recalibrate.
Waste pump malfunction.
Contact Technical Support
Bubble Detector
malfunction.
Contact Technical Support
Air in Calibrant A
Calibrant A.
Replace reagent kit with a new one,
prime, and recalibrate.
Tubing from reagent
module is disconnected,
plugged, or crimped.
Reconnect or replace tubing.
Calibrant A pump is not
working properly.
Contact Technical Support
15 MAINTENANCE PROGRAM
In order to ensure optimal performance and maximum useful life of the Autoanalyzer, it is important to
follow the cleaning and maintenance instructions outlined in this section.
Daily Maintenance
Before starting daily operation
1. Verify that the levels of the waste and washing solution containers are adequate to start with
the daily operation.
2. Run two diluter purge cycles, visually checking the absence of bubbles in the diluter body
Maintenance
Instrument
Diluter Purge
3. Run one wash cycle of the first 10 cuvettes, checking the absence of bubbles, and overflow of