Operator’s Manual: HTI BioChem FC-360
Page 62 of 217
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
Figure 7-29
4. Select the execution options: by
Selection,
by
Patients
or by
Reagents:
Selection
: This is the fastest option. The software organizes and establishes the work sequence, ordering
of the methods and samples, optimizing processing speed.
Patients
: Processes samples in the order in which they were requested.
Reagents
: All tests of the same method are run before it continues with the tests for the next method. The
user cannot select the methods’ processing order. The software establishes the optimal processing order.
Processing is grouped as
New Group
by default.
5. Press
Send
to continue or
Cancel send
if the operation is not carried out.
6. Select the methods for reagent verification on the
Reagents Verification
screen (Figure 7-30)
After sending samples to be processed, the
Reagent Verification
screen is displayed. In this screen, all
methods where the verification time (configured in the
Verification Time
field) has elapsed are indicated
with a check mark. It is recommended that the selected reagents are verified to ensure the integrity of the
reagents before processing samples.
Methods where the verification time has not elapsed can also be verified by selecting them on this screen.
It is recommended that methods are verified after changes have been made. Some examples of changes
that require verification are: when a new reagent is placed, when configuration of certain method
parameter is modified, etc.
Figure 7-30