Operator’s Manual: HTI BioChem FC-360
Page 34 of 217
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
Figure 6-8
Once all reagents have been positioned, type a name within the
Tray File
field and press
Save.
An
indefinite number of tray files may be created, each with difference names, and then press
Save
.
To load a stored reagent tray configuration, select from stored trays displayed in
Tray File
and then press
Load.
The tray file will be loaded and new reagent positions will appear on screen.
To delete a reagent tray, it must first be loaded by means of the
Load
key and then deleted by means of
the
Delete
key.
Figure 6-9
When the volume in a reagent bottle is low, an error message “lack of reagent” will be displayed indicating
which reagent(s) must be reloaded. This error windows has to be closed after reloaded process
Figure 6-10
When a lack of reagent message is displayed, the working sequence must be paused. Press
Pause
for a
few seconds until the instrument is paused.