Drift Error Flag Multiple
electrodes
May occur when new
electrode or Reagent Kit is
installed on system.
Purge Calibrant A and recalibrate.
Reference electrode.
Replace reference electrode and
recalibrate.
Electrical spike from
environmental source.
Contact Technical Support
Contact Technical Support
Component failure on ISE
Module board.
Contact Technical Support
Air in Sample
Insufficient sample pipetted
into ISE Module sample
entry port.
Probe with some blocked
Bubbles in Sample.
Sample must be free of bubbles.
Fluid leaks
Contact Technical Support
Sample not positioned.
Electrodes not seated properly. Remove
electrodes. Inspect O-rings and
reassemble.
Replace pump tubing.
Pump tubing obstructed.
Replace pump tubing.
Air in Sample and Calibrant A
Sample and Calibrant A
are segmented with air.
Electrodes are not properly seated or
compressed. Check compression tray,
spring, and seal. Remove and
reassemble electrodes.
Fibrin or salt is plugging
the electrode flow path.
Use Cleaning procedure.
Bubble detector is
malfunctioning.
Remove electrodes and clean or replace
electrode with plugged flow path. Reinstall
electrodes and recalibrate.
Contact Technical Support
Waste pump is
malfunctioning.
Contact Technical Support
Air in Calibrant B and Air in
Calibrant A
Calibrant B and Calibrant A
are segmented with air.
Electrodes are not properly seated. Check
compression tray, spring and seal.