12.8.5.2
Calibrant A purge
During this cycle, a volume of Calibrant A is pumped through the ISE module. This procedure eliminates
bubbles from the tubing.
Figure 12-53
12.8.5.3
Calibrant B purge
During this cycle, a volume of Calibrant B is pumped through the ISE module. This procedure eliminates
bubbles from the tubing.
Figure 12-54
12.8.5.4
Calibration
This cycle is used to calibrate the electrodes of the ISE Module. The ISE Module then cycles Calibrant B
and Calibrant A solutions in front of the electrodes and measures the millivolt output of the electrodes for
each of the respective solutions.
The ISE Module Calibration Cycle performs two successive calibrations.
Figure 12-55