Operator’s Manual: HTI BioChem FC-360
Page 212 of 217
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
(0386) Cuvettes Calibration
error (low energy in
Reference Beam)
Exit lens of the reference channel.
Contact Technical Support.
(0387) Cuvettes Calibration
error (Cuvette absorbance
too high, replace the
cuvettes)
Scratched cuvette or poorly aligned.
Replace the cuvette.(it is recommended
to replace the defective cuvettes)
Contact Technical Support.
Lack of calibration
Run photometer calibration
(0388) Cuvettes Calibration
error (Cuvette absorbance
too low, Calibrate Photometer
in a different cuvette or
replace de cuvettes)
Lack of calibration
Do photometer calibration
Cuvette not aligned.
Replace the cuvette.(it is recommended
to replace the defective cuvettes)
Contact Technical Support.
(0389) Cuvettes Calibration
(Cuvette absorbance too
high, replace the cuvettes
when possible)
Striped cuvette or misaligned
Check the alignment of the beam with the
cuvette.
Replace the cuvette.(it is recommended
to replace the defective cuvettes)
Lack of calibration
Run photometer calibration
(0390) Cuvettes Calibration
(Cuvette absorbance too low,
Calibrate Photometer in a
different cuvette or replace de
cuvettes when possible)
Lack of calibration
Run photometer calibration
Cuvette poorly aligned.
Replace the cuvette.(it is recommended
to replace the defective cuvettes)
Contact Technical Support.
(0391) Cuvettes Calibration
error (EEPROM writing error)
Defective EEPROM.
Contact Technical Support.
HEATER AND PREHEATER
Message ID Number and
Description
Probable cause
Possible Solution
(0401) Temperature reading
error (can't read Reaction
Tray A/D converter)
Wiring failure
Contact Technical Support.
A/D defective
(0451) Temperature reading
error (can't read Preheater
A/D converter)
Wiring failure
Contact Technical Support.
A/D defective
Contact Technical Support.