Operator’s Manual: HTI BioChem FC-360
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
Page 199 of 222
Message of a re-diluted result
or re-dilute manually with
results within the linear limit
Linearity configured incorrectly
Configure the linearity correctly: a
value between 0.80 y 0.95
Initial consumption configured
incorrectly
Configure the method’s initial
consumption correctly
Instrumental Problems
Instrumental problems may appear with visible failures in its components or by malfunction errors that are
shown up in the software. As well, instrumental problems may often be inferred from the analysis of observed
chemical problems.
Based on these observations, instrumental problems can be grouped according to the affected module or
component of the instrument. Decide which of the following components is probably affected and run the
associated checks and actions in each case:
-
Connections and supplies
-
Hydraulic system
-
Washer
-
Diluter
-
Sample/Reagent probe
-
Reaction cuvette tray
-
Sample/Reagent tray
-
Software
If this does not solve the problem, contact Technical Support
Connections and supplies
Observation
Probable cause
Action
Failure in the startup of the
instrument
Instrument disconnected
Plug the cable into the power source
Burnt out start up fuse
Replace the fuses
No electricity from the power source
Reestablish the electricity
Damaged start-up button
Contact Technical Support
Failure in the PC- instrument
communication
Instrument turned off
Turn on the instrument
Instrument turned off and on without
initialization procedure
Run initialization procedure
Invalid communication port
Close all open applications that might
be using the communication port and
re- open the software
PC disconnected from the instrument
Connect the PC-instrument
communication cable
17.3.1
Hydraulic system