Operator’s Manual: HTI BioChem FC-360
Page 24 of 217
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
Optical System
Light source:
a 12-Volt 20-Watt krypton-halogen lamp is used which has a high emission in the UV
range (320 nm-380 nm)
Collimating lenses:
3 plane-convex lenses are used.
Wavelengths:
340-380-405-450-505-546-578-600-650 and 700 nm and an extra filter as optional.
Beam splitter:
Divides the primary beam into two secondary beams: one goes towards the sample
channel and the other goes to the reference channel.
Sample channel:
The sample channel beam passes through the cuvette to the sample photosensor,
where the signal is read.
Reference channel:
The reference channel beam is read by the reference photosensor. The reference
channel compensates possible light source fluctuations.
Photometric Graph
Figure 4-6
1.
Lamp
2.
Plane-convex lens
3.
Stopper
4.
Filter wheel
5.
Beam splitter
6.
Sample cuvette
7.
Sample photosensor
8.
Reference photosensor