High Technology, Inc.
109 Production Road,
Walpole MA 02081 USA
Document Number
OM-E-FC-360-30
Revision Level
Rev. 3
Prepared by
J.S. Bolio
Approved by
S. Titov
Effective Date
8/15/2016
Operator’s Manual: HTI BioChem FC-360
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
Page 125 of 222
Access Control Tab
Access control can be changed by the software administrator.
Admin:
password to enter in Settings of Maintenance menu.
Methods:
password to enter in Settings of Methods, Calibrators and Controls menu.
Service:
password to enter in Commands Console of Maintenance Menu
Select on the left side the option that one to be changed and the right side fill in with the new password,
repeat new password and finally press
Change Password
button.
Figure 12-17
Warning: in case of losing or forgetting passwords contact technical support
Cuvettes Status
By clicking on
Maintenance
and selecting
Cuvettes Status
, a screen will appear (Figure 12-17) which
shows the number and status of each cuvette, represented by different colors:
Green: Clean Light Red: Dirty
Dark Red (6 different shades): In washing process and drying process:
-Washing 1: Aspiration of liquid and dispensing of washing solution
-Washing 2: Aspiration of liquid and dispensing of washing solution