High Technology, Inc.
109 Production Road,
Walpole MA USA 02081
Document Number
OM-E-FC-360-30
Revision Level
Rev. 3
Prepared By
J.S. Bolio
Approved By
S. Titov
Effective date
8/15/2016
Operator’s Manual: HTI BioChem FC-360
Page 210 of 217
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
Contact Technical Support.
Exit lens of the reference channel.
Contact Technical Support.
Amplifier sample Failure.
Contact Technical Support.
(0356) Photometer Calibration
error (low energy in Reference
Beam)
Exit lens of the reference channel.
Contact Technical Support.
Failure in the amplifier reference.
Contact Technical Support.
(0357) Photometer Calibration
error (low energy in Sample
and References Beams)
Defective lamp
Contact Technical Support.
Exit lens.
Contact Technical Support.
Incorrect positioning of a Beam Filter.
Contact Technical Support.
Defective filter
Contact Technical Support.
(0358) Photometer Calibration
error (high
Alignment failure in the optical path
Contact Technical Support.
energy in Sample Beam)
Sample Amplifier Failure.
Contact Technical Support.
(0359) Photometer Calibration
error (high energy in
Reference Beam)
Reference Amplifier Failure.
Contact Technical Support.
(0360) Photometer Calibration
error (high energy in Sample
and Reference Beams)
Incorrect positioning of a Beam Filter.
Contact Technical Support.
Lamp failure
Contact Technical Support.
Defective filter
Contact Technical Support.
Amplifiers failure
Contact Technical Support.
(0361) Photometer Calibration
error (high Zero in Sample
Amplifier)
Stray light.
Contact Technical Support.
Sample Amplifier Failure.
Contact Technical Support.
(0362) Photometer Calibration
error (high Zero in Reference
Amplifier)
Stray light.
Contact Technical Support.
Reference Amplifier Failure.
Contact Technical Support.
(0363) Photometer Calibration
error (high Zero in Sample
and Reference Amplifiers)
Incorrect positioning of a Beam Filter.
Contact Technical Support.
Lamp failure
Contact Technical Support.
Defective filter
Contact Technical Support.
Missing filter
Contact Technical Support.
Amplifiers failures
Contact Technical Support.