High Technology, Inc.
109 Production Road,
Walpole MA 02081 USA
Document Number
OM-E-FC-360-30
Revision Level
Rev. 3
Prepared by
J.S. Bolio
Approved by
S. Titov
Effective Date
8/15/2016
Operator’s Manual: HTI BioChem FC-360
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
Page 7 of 222
Figure 1-5
Once the screws have been removed and the instrument is released, the protective Styrofoam is removed
and the instrument can be placed on the working surface.
2
INSTALLATION
Installation must be carried out by trained personnel.
Electrical requirements
A standard 100 Volts 60 Hz or 220 Volts 50 Hz socket for 400 VA consumption. The supply line has a third
terminal to ensure proper ground connection. If line variations are higher than 10%, a ferroresonant
voltage regulator, or a 1500 VA uninterruptible power supply (UPS sine wave output) is recommended.
Dimensions & Weight
Width: 800 mm, Height: 464 mm, Depth: 647 mm
Weight: 48kg