High Technology, Inc.
109 Production Road,
Walpole MA USA 02081
Document Number
OM-E-FC-360-30
Revision Level
Rev. 3
Prepared By
J.S. Bolio
Approved By
S. Titov
Effective date
8/15/2016
Operator’s Manual: HTI BioChem FC-360
Page 178 of 217
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
Figure 15-1
ISE Module Maintenance Program (Optional)
Recommended Component Replacement Schedule (low volume user)
Li+ Electrode
Pump Tubing
Na+ Electrode
K+ Electrode
Cl- Electrode
Reference Electrode
Fluidic Tubing: 12 months
Recommended Replacement Schedule (high volume user)
(Greater than 100 samples per day)
Li+ Electrode: 3,000 samples
Pump Tubing: 6 months
Na+ Electrode: 10,000 samples
K+ Electrode: 10,000 samples
Cl- Electrode: 10,000 samples