Containers Status
By clicking on
Maintenance
and selecting
Containers Status
, a screen will appear which shows the
volume of liquid in each container, expressed as a percentage of its total capacity.
When the level of waste liquid is above 90% and/or that of the washing solution is below 10%, a red
warning sign will automatically appear on this screen (Figure 12-18)
Figure 12-19
To make changes in level of liquid of the bottles, the instrument must not be working. Once the waste
bottle has been emptied, the instrument can work again. Once new washing solution has been added,
make five purge cycles of the diluter
(Maintenance
Instrument
Diluter Purge)
Then wash cuvettes from 1-10 five times
(Operations menu
wash cuvettes
Range
1-10
Wash)
Warning: The level of liquid is only refreshed when processes that require the aspiration or
dispensing of liquid are carried out.
Programmed washing
The
Programmed Washing
option appears on screen two minutes after the working sequence has
finished and all the results have been shown (Figure 12-19)
This process washes dirty cuvettes which have just been used and those which have not been fully
washed, being executed in the indicated remaining time. If
Execute
is selected, the washing is carried out
at that moment.
If it were left some samples pending of process, passed 10 minutes an autostop will take place
automatically; after that the sequence of programmed washing will be carried out.
Warning: DO NOT cancel the programmed washing since it may affect next washings as well as