Operator’s Manual: HTI BioChem FC-360
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
Page 215 of 222
ISE
Message ID Number and
Description
Probable cause
Possible Solution
(0701) mV Out in
Serum/Plasma
Salt is plugging the electrode flow
path.
Remove electrodes and clean or
replace electrode with plugged flow
path. Reinstall electrodes and
recalibrate.
If single electrode is flagged
Replace problem electrode and
recalibrate.
If multiple electrode are flagged
Replace reference electrode and
recalibrate.
Electrical noise spike from
environmental source.
Contact Technical Support.
(0702) mV Out in Calibrant A
Salt is plugging the electrode flow
path.
Remove electrodes and clean or
replace electrode with plugged flow
path. Reinstall electrodes and
recalibrate.
If single electrode is flagged: May
occur when new electrode or
Calibrant A are installed.
Purge Calibrant A and recalibrate the
ISE module. If the electrode is new it
may initially drift as it rehydrates over
the course of 15 minutes.
If single electrode is flagged
Replace the electrode and recalibrate.
If multiple electrode are flagged
May occur when new electrode or
Reagent Kit is installed.
Purge the Calibrant A and recalibrate
the ISE module.
If multiple electrode are flagged
Reference electrode.
Replace reference electrode and
recalibrate.
(0703) Noise in
Serum/Plasma
If single electrode is flagged
Replace problem electrode and
recalibrate.
If multiple electrode are flagged
Replace reference electrode and
recalibrate.
Electrical noise spike from
environmental source.
Contact Technical Support.
Component failure on ISE Module
board.
Contact Technical Support
(0704) Noise in Calibrant A
If single electrode is flagged
Replace problem electrode and
recalibrate.
If multiple electrode are flagged
Replace reference electrode and
recalibrate.