C = ΔA/min • F
Bichromatic Reading
A
1
= 1
st
Absorbance Reading of the Sample
A
2
= 2
nd
Absorbance Reading of the Sample
A
N
= n
th
Absorbance Reading of the Sample
ΔA/min = change per minute of the sample absorbance calculated by linear regression over “n”
Absorbance readings
A
1
= 1
st
Bichromatic Absorbance Reading of the Sample
A
2
= 2
nd
Bichromatic Absorbance reading of the Sample
A
N
= n
th
Absorbance reading of the Sample
ΔA/min = change per minute of the sample bichromatic absorbance calculated by linear regression over
“n” Absorbance readings
F =
Factor programmed or obtained by the method calibration
C =
Concentration
C =
(ΔA/min - ΔA’/min) • F
8.1.4.4 Fixed Time Kinetics
In this type of measurement the instrument makes only two readings, and calculates the abs.
between both for the end result (Figure 8-10). This type of measurement is used for kinetic methods
such as Urea, Creatinine, etc.
Figure 8-2