Operator’s Manual: HTI BioChem FC-360
Page 8 of 217
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
Environmental requirements
Operating temperature: 20°C ~25°C
Operating Humidity: 20~85% (no condensation)
Storage Temperature: 0°C~50°C
Stora
ge Humidity: ≤80%
Moving the instrument
Avoid hitting the instrument as well as subjecting it to any kind of vibration while moving it.
PC minimum requirements
2 GHz Pentium IV
RAM Memory, 512 MB
Independent video card
40 GB Hard Drive
CD Rom
CD Reader
Windows XP
Printer (optional)
The PC must comply with safety regulations IEC 60950 and have a certified source.
Instrument placement and preparation
Warning: Avoid hitting the instrument as well as subjecting it to any kind of vibration while moving
it.
The instrument must be placed on a counter in a room free from dust and corrosive vapors to ensure its
proper operation.
The instrument must not be subject to vibrations or sudden changes of temperature.
Leave a 20 centimeter separation between the equipment and the wall to ensure proper ventilation. Liquid
containers must be under the counter. Avoid collapsing and/or curving of waste exit tubing.
Avoid any centrifuge, lifts or x-ray supply lines or any other kind of noise-generating equipment in the
supply line.
Avoid direct sun light or illumination on the instrument.
The supply line has a third ground conductor for protection purposes.