Operator’s Manual: HTI BioChem FC-360
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
Page 69 of 222
Stop
The instrument can be completely stopped by pressing
Stop
on the
Operations Menu
or the
Stop
icon
(Figure 7-36). The processing of samples and the reading of reagents in process will stop completely as a
result.
Figure 7-42
Stop
is used:
When there is a mechanical error. For example, if there is a problem with the probe, cuvettes tray
cover is out of place, washing solution is needed, etc.
When there is an error in the processing. For example: a reagent has been positioned in the
wrong place, samples are missing or they have been incorrectly placed, reagent deterioration
index is out of range, etc.
When there is an error sign on the screen that indicates Retry or Stop the process.
When the instrument is running tests and the option
Stop
is pressed, a warning sign will appear (Figure 7-
37) to confirm the operation by pressing
Yes
or
No
.
Figure 7-38
Pause
The instrument can be paused by pressing
Pause
on the
Operations Menu
or the
Pause
icon (Figure 7-
38). Dispensing of samples will stop but reactions in process will continue, including R2 dispensing.