Shutdown Procedure
Preparing the ISE Module for Storage if the laboratory plans to store the ISE Module, the following steps
should be performed:
Before removing the electrodes, they should be cleaned using the cleaning solution and then 3 Purge A
cycles should be run.
Enter the Maintenance Cycle of the analyzer to purge the ISE Module fluid path. Maintenance
Instrument
ISE
Electrode replacement.
Depress the compression tray and remove all electrodes, including the reference electrode from
the ISE Module.
Place the Na+ and Cl- electrodes into individual sealed bags.
Reinsert the Reference Electrode flow path line with yellow flag, if available, and then put into
individual sealed bags.
Aspirate a small volume of Calibrant A from the top port of the reagent kit into a syringe fitted
with a blunt needle.
Inject sufficient volume of Calibrant A into the lumen of the K+ and Li+ electrodes until fluid fills
the lumen.
Cover both ends of the lumen (both sides of the K+ and Li+ electrodes) with tape to hold the
Calibrant A in place.
Insert the K+ and Li+ electrodes into a sealed bag.
Remove the reagent kit from the analyzer and discard.
Remove all fluidic tubing and thoroughly rinse with DI water.
ISE Module re-activation
Remove all electrodes from sealed bags.
Remove tape from K+ and Li+ electrode.
If necessary, soak the reference electrode in warm water until the lumen of the electrode has
been cleared of salt build-up.
Install electrodes into the ISE Module.
Connect new reagent kit to the ISE Module.
Use Reagent Kit Replacement to prime the calibrants. Maintenance
Instrument
ISE
Reagent Kit Replacement
Calibrate ISE module. Maintenance
Instrument
ISE
Calibrate
ISE Troubleshooting
Overview
To enhance trouble-free operation of the ISE Module, it is important to follow the recommended
component replacement schedule listed in the maintenance section of this manual.