Operator’s Manual: HTI BioChem FC-360
Page 66 of 217
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
with a check mark. It is recommended that the selected reagents be verified to ensure the integrity of the
reagents before processing samples.
Methods where the verification time has not elapsed can also be verified by selecting them on this screen.
It is recommended that methods be verified after changes have been made. Some examples of changes
that require verification are: when a new reagent is placed, when configuration of certain method
parameter is modified, etc.
Figure 7-36
7. Press
Verify Selected
or
Do not verify
. If the option
Do Not Verify
is selected, the methods that are
checked will not be verified.
Warning: This is the last screen where processing of requested samples can be cancelled.
8. Press
Start
on the screen to put the samples in the tray (Figure 7-31)
Warning: It is not possible to cancel processing of selected samples in this step.