Saving Calibrators Files
Calibrators configuration data in use can be saved through the submenu
Save
. By pressing
Calibrators
Menu
Save
, an exploration window is displayed. Select the folder or location where the data will be
saved. Name the file and press
Save
or
Cancel
if the operation will not be carried out. A file with extension
*.acf will be created which can be used as a backup or to be installed in another PC.
Loading Calibrators Files
Calibrators configuration data can be obtained through the submenu
Load
from a different file than the file
in use. By pressing Calibrators
Menu
Load
, an exploration window is displayed. Select the folder or
location from where the data will be obtained. Select the file with extension *.acf and press
Open
or
Cancel
if the operation will not be carried out.
All the data of the calibrators that appear in the selected file will then be entered.
Warning: By entering configuration data of new calibrators through the submenu Load, all the data
of calibrators in use will be deleted. Additionally, if the methods configured in the loaded file are
different from the methods in use, the calibrator value will appear as -1.
Before this operation, back up the calibrators in use as indicated in Chapter 9 Point 9.2
10 CONTROLS MENU
Controls parameters setting as well as controls files handling can be carried out from the
Controls Menu.
There is also a shortcut to parameters setting from the icon
Controls Parameters
(Figure 10 - 1)
Figure 10-1