Operator’s Manual: HTI BioChem FC-360
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
Page 61 of 222
-Non-correlative samples:
Holding the Control key on the keyboard, select each sample with the mouse.
Either all or some of the requested samples can be selected; the remaining samples can be left for later
process.
Warning:
Verify that the methods have been calibrated and controls have been run to determine validation
of the calibration. After reagent blank values, calibrator absorbance and controls have been validated;
continue with the processing of patients.
Figure 7-27
2. Press the
Samples to Tray
icon. The
Samples Position
screen will be displayed (Figure 7-28), where
Type
of requested sample (patient, calibrator, control: PAT, CAL, CON),
Name, ID
and
Position
are
indicated.
3. Check that the actual positioning of samples on the tray corresponds to what appears on the
Samples
Position
screen. Sample positions can be modified by using the drop down window in the Position
column.
Figure 7-28
If the same position number is used by mistake for two different samples in the same run, an error
message will be displayed requesting to correct the repeated positions (Figure 7-29).