Failure to update a calibration
Calibration made with the method
setting screen open
Recalibrate with the method setting
screen closed or edit the value
obtained for the calibration
Error message: The instrument
cannot be initialized
Instrument turned off
Turn on the instrument
Instrument turned off and on without
initialization procedure
Run initialization procedure
Burnt out start up fuse
Replace the fuses
No electricity from the power source
Reestablish the electricity
Damaged Analyzer on/off button
Contact Technical Support
Instrument disconnected
Plug the cable into the power source
Invalid communication port
Close all open applications that might
be using the communication port and
re- open the software
PC disconnected from the instrument
Connect the PC-instrument
communication cable
Mechanical failure
Contact Technical Support
For any problem other than those described here, contact exclusively the Technical Support authorized by
HTI
Hardware Messages
Error messages will be displayed on the screen like this:
Figure 17-1
Figure 17-2