(0364) Photometer Calibration
error (EEPROM writing error)
Defective EEPROM
Contact Technical Support.
(0365) Photometer Calibration
(High gain Sample Amplifier)
Scratched cuvette or not aligned.
Replace the cuvette or try to calibrate
another cuvette. (Anyway, it is
recommended to replace defective
cuvettes)
Contact Technical Support.
Exit lens of the reference channel.
Contact Technical Support.
(0366) Photometer Calibration
(High gain Reference
Amplifier)
Exit lens of the reference channel.
Contact Technical Support.
(0367) Photometer Calibration
(High gain Sample and
Reference Amplifiers)
Defective lamp
Contact Technical Support.
Exit lens.
Incorrect positioning of a Beam
Filter.
Defective filter
CUVETTE CALIBRATION
Message ID Number and
Description
Probable cause
Possible Solution
(0381) Cuvettes Calibration
error (no filters programmed
in EEPROM)
Filter ID values set on the EEPROM
Contact Technical Support.
(0382) Cuvettes Calibration
error (Filter Wheel Controller
communication)
Possible failure in IC filter-wheel
controller.
Contact Technical Support.
(0383) Cuvettes Calibration
error (Amplifier Controller
communication)
Possible failure in IC controller
(Amplifier).
Contact Technical Support.
(0384) Cuvettes Calibration
error
Error in the logic of calibration.
Contact Technical Support.
(0385) Cuvettes Calibration
error (low energy in Sample
Beam)
Lack of calibration.
Run photometer calibration.
Scratched cuvette or poorly aligned.
Contact Technical Support.
Replace the cuvette.(it is recommended
to replace the defective cuvettes)
Exit lens of the sample channel
Contact Technical Support.
Lack of calibration.
Run photometer calibration.