Hydraulic System
Figure 4-5
Its main components are the liquid diaphragm pump, the solenoid valve and the diluter, the washing
station and the tubing that connects the wash solution container to the probe.
The diluter aspirates the reagent and a sample volume required, and dispenses them into the reaction
cuvette.
The liquid diaphragm pump conveys the wash solution from the wash solution container to the end of the
probe capillary. This allows the washing and purging of the tubing, the diluter and the probe.
It is important to highlight that reagents never reach the diluter’s body.