Operator’s Manual: HTI BioChem FC-360
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
Page 11 of 222
Figure 2-3
Warning: The tubing with the biggest diameter (waste funnel) must be kept straight downwards
without curves to allow free drain of fluids
.
Washing solution
Warning: Containers must be placed under the analyzer.
Warning: Container must not be placed on the same analyzer level.
2.7.1.1
Washing solution for probe and cuvettes
First
wash concentrate solution
has to be prepared from Triton X-100 Solution. Then, use wash
concentrate solution to prepare washing solution for probe and cuvettes.
Wash Concentrate
: dilute 1 part v/v of Triton X-100 in 9 parts of distilled water, Ex: 100 ml Triton X-
100 in 900 ml of distilled water.
The Triton X-100 is extremely viscous. It is suggested to pour the desired quantity into a graduated
cylinder and gradually add the water. Heat the distilled water to 70
– 80 ºC to help dissolve the Triton
X-100.
Washing solution for probe and cuvettes
: dilute 7 ml of wash concentrate in 1 liter of distilled water.
Ex: 140 ml of wash concentrate per 20 liters of distilled water.