15.6.9.1
Replacing ISE peristaltic pumps tubing (For analyzers with ISE Module installed)
Figure 15-15
1. Remove the reaction Tray cover.
2. Locate the front black plate supporting the three peristaltic pumps and remove the tubing connection
from the yellow connectors.
3. Remove the Peristaltic Pumps to release the Peristaltic Tubing.
16 PREPARING SOLUTIONS
Wash solution.
Warning: Containers must be placed under the level of the analyzer, never on the same level
Washing solution for probe and cuvettes
First,
Wash Concentrate Solution
(mother solution) has to be prepared from Solution of Triton X- 100,
this concentrate solution will be used to prepare washing solution for probe and cuvettes.
Wash Concentrate (mother solution)
: dilute 1 part v/v of Triton X-100 in 9 parts of distilled water, Ex:
100 ml Triton X-100 in 900 ml of distilled water.
The Triton X-100 is extremely viscous. It is suggested to pour the desired quantity into a graduated
cylinder and gradually add the water. Heat the distilled water to 70
– 80 ºC to help dissolve the Triton X-
100.
Washing solution for probe and cuvettes:
dilute 7 ml of the mother solution in 1 liter of distilled water.
Ex: 140 ml of wash concentrate for 20 liters of distilled water.