![HTI BioChem FC-360 Скачать руководство пользователя страница 148](http://html.mh-extra.com/html/hti/biochem-fc-360/biochem-fc-360_operators-manual_2179333148.webp)
High Technology, Inc.
109 Production Road,
Walpole MA USA 02081
Document Number
OM-E-FC-360-30
Revision Level
Rev. 3
Prepared By
J.S. Bolio
Approved By
S. Titov
Effective date
8/15/2016
Operator’s Manual: HTI BioChem FC-360
Page 140 of 217
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
The P150 CV% should be equal to or less than 1.5% and The P250 CV% should be equal to or less
than 2.5%
The following is an example of a precision study
P150
P250
1
0.9866
1
1.7018
2
0.9950
2
1.7215
3
0.9906
3
1.7327
4
0.9776
4
1.7220
5
0.9775
5
1.7476
6
0.9741
6
1.7450
7
0.9824
7
1.7105
8
0.9770
8
1.7218
9
0.9798
9
1.7087
10
0.9827
10
1.7184
MEAN
0.9823
MEAN
1.7230
SD
0.01
SD
0.01
CV%
0.67
CV%
0.87
CV%
Range
≤ 1.5
CV%
Range
≤ 2.5