Operator’s Manual: HTI BioChem FC-360
Page 78 of 217
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
End Point Reagent Blank
End Point Sample Blank
Kinetics
Fixed Time Kinetics
ISE (Parameters at the end of this Chapter) (Optional)
Main Wavelength:
Enter the main wavelength for the proper reaction, as provided by the reagent
manufacturer. The software will select the filter that is closer to that value.
Bichromatic Wavelength:
Enter the bichromatic wavelength for the proper reaction, as provided by the
reagent manufacturer. The software will select the filter that comes closest to that value. For
monochromatic mode, enter “0”.
Unit:
Enter the units of measurement desired for the calculation and printout of the results.
Decimals:
Enter the number of decimal points desired for the calculation and printout of the results.
Sample Volume:
Enter the sample volume according to the reagent/sample ratio provided by the reagent
manufacturer. The volume sample should not be less than 2 µl.
R1 Volume:
Enter the volume of reagent 1 according to the reagent/sample ratio provided by the reagent
manufacturer.
R2 Volume
: Enter the volume of reagent 2 according to the reagent 1/reagent 2/sample ratio given by the
reagent manufacturer if it corresponds. Leave it at “0” for single reagent methods.
Warning: The total volume (R1+ R2 + sample) must be less than 500 µl. The minimum reading
volume should be 220 µl. The maximum cuvette capacity is 600 µl.
Time to Dispense Reagent 2:
Enter the time to dispense the second reagent after the dispensing of
reagent 1 and the sample. If 0 sec is entered the R1 + R2 will be dispensed at the same time.
Reagent Deterioration Index:
The instrument checks the state of the reagents according to the reagent
manufacturer’s instructions. This value corresponds to the reagent blank’s absorbance. If the measured
absorbance value is outside the programmed range, a “deteriorated reagent” message will be shown in the
operations tab beside each result.
Min.:
Enter the minimum absorbance value for the reagent blank according to the reagent manufacturer’s
instructions. If a minimum absorbance is not given, enter 0 as the value.
Note: With some bichromatic methods, a negative result is obtained for the reagent blank; in this
case enter an absorbance value less than 0.
Max.:
Enter the maximum absorbance value for the reagent blank as indicated by the reagent
manufacturer. If a maximum absorbance is not given, enter 2 as the value.