Operator’s Manual: HTI BioChem FC-360
Page 152 of 217
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
Figure 12-59
12.8.5.9
Reagent Kit Replacement
Reagent kit should be replaced any time the system asks to, when controls do not fall within the range
and/or when the calibration of the module shows error.
A pop up window will appear, replace the reagent kit and press Accept.
Figure 12-60
The system will start purging to eliminate bubbles from the tubing.
Place a tube with saline solution in the required position, and then press Ok. This cycle is used to
calibrate the peristaltic pumps of the ISE Module.