High Technology, Inc.
109 Production Road,
Walpole MA 02081 USA
Document Number
OM-E-FC-360-30
Revision Level
Rev. 3
Prepared by
J.S. Bolio
Approved by
S. Titov
Effective Date
8/15/2016
Operator’s Manual: HTI BioChem FC-360
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
Page 205 of 222
Message ID Number and
Description
Probable cause
Possible Solution
(0151) Probe impact
(Horizontal Movement)
Physical impact
Verify possible obstacles in the
movement
Impact switches failures (preheater
board)
Contact Technical Support.
Poor contacts
Contact Technical Support.
(0152) Probe Horizontal
Movement not initialized or
initialization error (Home
active)
Home Sensing failure
Contact Technical Support.
Mechanical failure
Contact Technical Support.
Horizontal movement not initialized
Execute the command to initialize.
(0153) Probe Horizontal
Movement not initialized or
initialization error (Home
inactive)
Home sensing failure
Contact Technical Support.
Mechanical failure
Contact Technical Support.
Horizontal movement not initialized
Execute the command to initialize.
(0154)
Probe
Horizontal
Movement error (Home active)
Home sensing failure
Contact Technical Support.
Mechanical failure
Contact Technical Support.
(0155)
Probe
Horizontal
Movement
error
(Home
inactive)
Home sensing failure
Contact Technical Support.
Mechanical failure
Contact Technical Support.
(0156) Upgrade Horizontal
Movement Controller
Old Controller version.
Contact Technical Support.
REACTION TRAY
Message ID Number and
Description
Probable cause
Possible Solution
(0201) Probe Impact
(Reaction Tray)
Physical impact
Verify possible obstacles in the
movement.
Impact switches failure (preheater
board)
Contact Technical Support.
Poor contacts
Contact Technical Support.
(0202) Reaction Tray not
initialized or initialization error
(Home active)
Home sensing failure
Contact Technical Support.
Mechanical failure
Contact Technical Support.
Reaction tray not initialized
Execute the command to initialize.
(0203) Reaction Tray not
initialized or initialization error
(Home inactive)
Home sensing failure.
Contact Technical Support.
Mechanical failure
Contact Technical Support.
Reaction tray not initialized.
Execute the command to initialize.
(0204) Reaction Tray not
initialized
Module not initialized.
Execute the command to initialize.