Operator’s Manual: HTI BioChem FC-360
Page 204 of 217
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
(0037) EEPROM reading error
(variable does not exist)
Wrong Direction
Contact Technical Support.
VERTICAL
Message ID Number and
Description
Probable cause
Possible Solution
(0101) Probe Impact (Vertical
Movement)
Physical impact
Verify possible obstacles along the
path of header movement.
Impact switches failures (preheater
board)
Contact Technical Support.
Poor contacts
Contact Technical Support.
(0102)Probe Vertical
Movement not initialized or
initialization error (Home active)
Home sensing Failure
Contact Technical Support.
Mechanical failure
Contact Technical Support.
Vertical movement not initialized.
Execute the commands to initialize the
movement
(0103)Probe Vertical
Movement not initialized or
initialization error (Home
inactive)
Sensing failure home
Contact Technical Support.
Mechanical failure
Contact Technical Support.
Vertical movement not initialized
Execute the commands to initialize the
movement
(0104) Probe Vertical
Movement error (Home not
found)
Home sensing failure
Contact Technical Support.
Mechanical failure.
Contact Technical Support.
(0105) Probe vertical movement
error (Home active)
Home sensing failure
Contact Technical Support.
Mechanical failure
Contact Technical Support.
(0106) Upgrade Vertical
Movement Controller
Controller old version.
Contact Technical Support.
(0107) Probe level not
detected(level inactive)
Wiring failure
Contact Technical Support.
Probe failure.
Contact Technical Support.
Preheater PCB failure
Contact Technical Support.
(0108) Probe Impact (Vertical
Movement) and incorrect Probe
Level Detection (Level active)
See error solving 101 and 109.
(0109) Incorrect Probe Level
Detection (Level active)
Sensing failure
Clean the probe in case of possible
dirtiness or a drop between the
auxiliary and the main capillary.
Probe failure
Contact Technical Support.
Module preheater failure
Contact Technical Support.
Wiring failure
Contact Technical Support.
HORIZONTAL