Messages
Message
Location
Explanation
What to do
●
Headbox off
●
param.
Headbox is not connected.
●
Connect the headbox.
●
High EMG
●
param.
High frontal muscle activity may
disturb the EEG measurement.
●
Calm the patient.
●
Identical EEG
modules
●
al. area
There are two or more EEG
modules in the system.
●
Remove all but one EEG
module.
●
Poor electr. contact
●
param.
Electrode impedance is over 5
kOhm.
●
Press the electrode to
improve connection.
●
Change the electrode.
●
Check the impedance.
Messages related to BIS measurement
For information regarding alarm priorities and escalation times, see the CARESCAPE Modular
Monitors Supplemental Information Manual.
Make sure you are familiar with the generic layout of the screen. This will help you identify
where on screen the following messages appear. The message location is indicated with the
following abbreviations:
●
al. area = alarm area
●
param. = parameter window
●
report = report view
●
wavef. = waveform area
Message
Location
Explanation
What to do
●
Apply sensor
●
param.
The sensor connection to the
patient may be loose.
●
Press the BIS sensor
electrodes to improve
connection.
●
Artifact
●
param.,
wavef.
Signals contain noise or artifact.
●
Check sensor contact.
●
Remove sources of excessive
noise.
●
Automatic check off
●
param.
Automatic sensor check has
been turned off.
●
If required, activate the
automatic check.
●
BIS cable off
●
al. area
Cable is disconnected from the
module.
●
Connect the cable.
●
BIS DSC error
●
al. area
The BISx is not communicating
or operating properly.
●
Test the BISx.
●
If the problem persists,
contact authorized service
personnel.
●
BIS high
/
BIS low
●
al. area
Measurement values are equal
to or outside the alarm limits.
●
Check the patient status.
●
Check the dosage of
anesthetics.
●
Adjust alarm limits if
necessary.
2062971-001
CARESCAPE Modular Monitors
501
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...