ECG
Setting alarm limits for individual leads
1.
Select the ST parameter window.
2.
Select
Alarms
.
3.
Select
Alarm On
for an ECG lead to adjust its alarm limits.
If the alarm is locked, there is a lock symbol beside the check box and the
selection is not available.
4.
Set
High
and
Low
alarm limits with the arrows.
Setting relative alarm limits
You can adjust the high/low alarm limits set around the current ST value for all of the
individual ST leads or for the leads in a selected lead group. For example, when you
select a
Relative Auto Limits
value of 2 mm, the high limit is set at the current ST value
+2 mm, and the low limit is set at the current ST value -2mm.
1.
Select the ST parameter window.
2.
Select
Alarms
.
3.
Select
Relative Auto Limits
.
4.
Set the relative limits as needed:
●
Set
All Limits
with the arrows and select
Update All
.
●
Set limits for
Anterior
leads with the arrows and select
Update Anterior
.
●
Set limits for
Inferior
leads with the arrows and select
Update Inferior
.
●
Set limits for
Lateral
leads with the arrows and select
Update Lateral
.
QT detection
This selection is not available with combination monitoring or TRAM.
With Multi-lead QT/QTc Analysis license only.
The administration of some drug types can prolong the QT segment. Monitoring QT
segment changes can help identify how these drugs are affecting the QT segment.
QT/QTc measurement limitations
●
Not available with combination monitoring or TRAM.
●
At least one measured V-lead must be available in order for the algorithm to
process QT.
●
QT/QTc values are calculated with 5-leadwire, 6-leadwire, or 10-leadwire ECG
cables.
Starting the QT/QTc measurement
This selection is not available with combination monitoring or TRAM.
1.
Select the HR parameter window.
2.
Select the
QT
tab.
3.
Select
On
from the
QT Analysis
list.
Setting QT/QTc alarms
This selection is not available with combination monitoring or TRAM.
2062971-001
CARESCAPE Modular Monitors
189
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...