Invasive pressures
GE recommends that the signal source used to trigger an IABP should be the
intra-aortic balloon pump itself. This ensures that the trigger signal is compatible with
all modes of the IABP. An extra set of ECG electrodes or an additional connection
from the arterial line can be connected to the monitor to produce waveforms on the
monitor’s display for consolidated viewing.
When using the monitor for triggering, the IABP triggers off the PDM or TRAM module’s
Defib Sync ports and uses the first zeroed arterial invasive pressure channel. If the
intra-aortic balloon pump triggers off arterial pressure, the analog output defaults to
the numerically first zeroed arterial pressure:
Art
,
Fem
, or
UAC
. If no arterial pressure
is available, the numerically first zeroed pressure is used.
Triggering intra-aortic balloon pumps
NOTE
If you choose to use the monitor for triggering, use the
following instructions. Failure to follow these instructions may
result in an incompatible analog output signal, which may
contribute to patient injury.
1.
Contact the balloon pump manufacturer for interface requirements. See the
technical specifications for the ECG analog output delay specification for the
acquisition device.
2.
Connect a compatible analog output cable to the monitor through the PDM or
TRAM module’s Defib Sync connector.
3.
Adjust the invasive pressure filter. If arterial pressure is used to trigger the balloon
pump, use the 40 Hz pressure filter.
4.
Primary displayed ECG lead: If the balloon pump triggers off the R wave of the
ECG, review the patient’s ECG leads and place the one with the greatest amplitude
in the top (primary) position on the monitor display.
5.
Pacemaker detection: If the patient has a pacemaker, be sure the pacemaker
detection is turned on. Failure to turn pacemaker detection on may cause poor
beat detection, which may result in inadequate triggering of the balloon pump.
Effects of IABP on displayed values
Displayed pressure values are affected by the intra-aortic balloon pump. The IABP
program displays three values, for example 150/45 (98). The first value is the systolic
value, the second is the diastolic value, and the third is the mean value. The displayed
numeric values are indicating a rapidly varying waveform generated during IABP
treatment and do not always reflect a true arterial blood pressure.
For accuracy and reliability, always combine two or more of the recommended
methods for reading arterial and/or femoral blood pressure:
●
The IABP waveform displayed on the screen (use scales for evaluation),
●
A printed copy of the waveform (use scales for evaluation),
●
The display on the balloon pump device, if available.
Since there are a number of points along the IABP waveform that could be the
displayed value, it is important to know which points the program uses. The values
displayed will differ depending on the timing of the pump.
1:1 or 1:2 timing
Diastolic numerics: The displayed diastole always equals the balloon end diastole.
250
CARESCAPE Modular Monitors
2062971-001
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...