Temperature
●
For more detailed information regarding the temperature probes, refer to their
own instructions for use.
●
Depending on the temperature module used, not all temperature measurements
and settings are available to view or change.
●
The temperature measurement uses direct mode. Displayed temperature values
represent the probe temperature of the measurement site on the patient.
●
If two temperature measurement modules that map to the same temperature
channel are connected, the first detected temperature measurement module is
assigned to the indicated temperature channel. When this occurs, a technical
alarm indicating identical modules is generated.
●
A temperature channel is activated when the module detects a temperature probe.
●
A temperature channel is deactivated when a temperature probe is detached
from the module.
●
A deactivated temperature channel does not release the assigned temperature
channel. To release an assigned temperature channel, remove the module from
the system.
Temperature measurement setup
Temperature equipment to patient connection
1.
Module with temperature measurement capability
2.
Dual temperature cable
3.
Single temperature cable
4.
Temperature interconnect cable for disposable temperature probes
5.
Reusable temperature probe
6.
Disposable temperature probe
Preparing the patient for temperature measurement
1.
Follow the manufacturer's instructions for probe application and instructions.
2.
Connect the single or dual temperature cable to the acquisition module connector.
256
CARESCAPE Modular Monitors
2062971-001
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...