Alarms
The lower the number, the quieter the alarm volume. Note that the minimum allowed
alarm volume levels are set in the
Care Unit Settings
.
Audible and visual alarm signals
Alarm signals indicate that an alarm condition is present. The alarm priority levels are
indicated by visual and audible signals. The visual and audible alarm signals assume
that the patient monitor and the operator are within the patient environment (1.5
meters).
When physiological alarm messages are received from other monitors on the S/5
Network, only the most recent and highest priority message is shown. There are no
flashing numerics or background color. No messages are shown in the parameter
window area.
The following table lists alarm signals for different alarm priority levels:
Priority level
Signal
High
Medium
Low
Informational
Parameter window
physiological data values
Black text flashes inside
a red box.
Black text flashes
inside a yellow box.
Black text inside a
cyan (blue) box.
Not applicable.
Alarm area
White text inside a red
box.
Black text inside a
yellow box.
Black text inside a
cyan (blue) box.
Black text inside a
grey box.
Waveform field messages
Text
Text
Text
Text
Audible tone pattern
1
(IEC
60601-1-8)
Repeats pattern of 2 * 5
beep tones
Repeats pattern of
3-beep tones
1-beep tone
None
Audible tone pattern
(legacy)
Repeats pattern of
3-beep tones (crisis)
Repeats pattern
of 2-beep tone
(warning)
1- beep tone
(advisory)
None
Alarm light indicator
2
Flashes red
Flashes yellow
Solid blue
No effect
Automatic view on alarm
audible tone pattern
User configurable
3
User configurable
3
User configurable
3
None
1
The IEC audible tone pattern is the factory default setting.
2
When the audible alarms are turned off or are paused, the alarm Audio pause/off area of the Alarm light is a solid
blue color.
3
The AVOA audible alarm tone pattern may be configured to match the local monitor’s audible tone pattern, repeat a
pattern of two beep tones, beep twice upon activation by an alarm condition, or be turned off. For configuring the AVOA
audible alarm tone pattern, see the CARESCAPE Modular Monitors Supplemental Information Manual.
Auditory information signals
The monitor performs a self-diagnostic procedure at start-up and generates an
auditory test signal. There are also other auditory information signals indicating the
status of some parameter measurements.
For more information, see the CARESCAPE Modular Monitors Supplemental
Information Manual.
Audible alarms off behavior
Depending on the
Audio Alarm
default settings configured during installation, you
can turn on or turn off audible alarms.
136
CARESCAPE Modular Monitors
2062971-001
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...